Model RFT9739 Field-Mount TransmitterInstruction ManualVersion 3 TransmittersFebruary 2000
viRFT9739 Field-Mount Transmitter Instruction ManualContents continuedFigure 7-1 Diagnostic LED and communicator loops . . . . . 64Figure 7-2 HART® Co
90RFT9739 Field-Mount Transmitter Instruction ManualTheory of Operation continuedPressure compensationA pressure transmitter can be connected to the R
RFT9739 Field-Mount Transmitter Instruction Manual91AppendixD HART® Communicator Menu TreesFigure D-1. On-line menu1 Mass flow2 Temperature3 Mass tota
92RFT9739 Field-Mount Transmitter Instruction ManualHART® Communicator Menu Trees continuedFigure D-1. On-line menu continued1 Process variables See p
RFT9739 Field-Mount Transmitter Instruction Manual93HART® Communicator Menu Trees continuedFast keyThe fast key code is a sequence of numerical button
94RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual95AppendixE Label Maintenance and ReplacementMaintaining and replacing labelsMicro Motion product sa
96RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual97AppendixF Transmitter Version IdentificationTo identify a Version 3 RFT9739 field-mount transmitte
98RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual99AppendixG Replacing Older TransmittersStep 1 Disconnecting the old transmitterFollow these steps
RFT9739 Field-Mount Transmitter Instruction Manual1Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sens
100RFT9739 Field-Mount Transmitter Instruction ManualReplacing Older Transmitters continuedStep 2 Determining type of RTD in the sensorDetermine whet
RFT9739 Field-Mount Transmitter Instruction Manual101Replacing Older Transmitters continuedStep 3 Installing the RFT9739 transmitterFollow these inst
102RFT9739 Field-Mount Transmitter Instruction ManualReplacing Older Transmitters continuedFigure G-2. RE-01 Remote Electronics Unit terminalsTable G
RFT9739 Field-Mount Transmitter Instruction Manual103Replacing Older Transmitters continuedFigure G-3. RFT9712 Remote Flow Transmitter terminalsTable
104RFT9739 Field-Mount Transmitter Instruction ManualReplacing Older Transmitters continuedStep 4 Characterizing sensors with copper RTDsUse the proc
RFT9739 Field-Mount Transmitter Instruction Manual105AppendixH Return PolicyGeneral guidelinesMicro Motion return procedures must be followed for you
106RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual107Page numbers in bold indicate illustrations.IndexAAbout this manual 1Accuracy 77Agency approval
108RFT9739 Field-Mount Transmitter Instruction ManualIndex continuedDD, DL, DT sensoraccuracy with 77normal resistance range 73repeatability with 7
RFT9739 Field-Mount Transmitter Instruction Manual109Index continuedGrounding 18–20, 20power supply and 18, 19HHandbook 44custody transfer event reg
2RFT9739 Field-Mount Transmitter Instruction ManualBefore You Begin continuedFigure 1-1. RFT9739 exploded viewHinged cover ofelectronics moduleIntrins
110RFT9739 Field-Mount Transmitter Instruction ManualIndex continuedmA 27–29Bell 202 multidrop network 48communication configuration 9–10performanc
RFT9739 Field-Mount Transmitter Instruction Manual111Index continuedSafety barrier partition 16general wiring requirements 15power-supply wiring 18
112RFT9739 Field-Mount Transmitter Instruction ManualIndex continued"RAM error" 68"RTI error" 68"security breach" 71–
recycled paperMicro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301Tel (303) 530-8400(800) 522-6277Fax (303) 530-84
RFT9739 Field-Mount Transmitter Instruction Manual3Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sens
4RFT9739 Field-Mount Transmitter Instruction ManualGetting Started continuedInstallations in EuropeTo comply with CENELEC standards for hazardous area
Getting Started continuedRFT9739 Field-Mount Transmitter Instruction Manual5Output Wiring Startup TroubleshootingBefore You Begin Getting Started Moun
6RFT9739 Field-Mount Transmitter Instruction ManualGetting Started continuedTable 2-1. Security modesSecurity mode 8When transmitter security is set f
Getting Started continuedRFT9739 Field-Mount Transmitter Instruction Manual7Output Wiring Startup TroubleshootingBefore You Begin Getting Started Moun
8RFT9739 Field-Mount Transmitter Instruction ManualGetting Started continuedTo make changes to configuration or calibration parameters once security m
Getting Started continuedRFT9739 Field-Mount Transmitter Instruction Manual9Output Wiring Startup TroubleshootingBefore You Begin Getting Started Moun
10RFT9739 Field-Mount Transmitter Instruction ManualGetting Started continuedTable 2-2. Communications configurationInstructionsBefore beginning, make
RFT9739 Field-Mount Transmitter Instruction Manual11Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sen
12RFT9739 Field-Mount Transmitter Instruction ManualTransmitter Mounting continued3.2 Mounting to a wallFollow these guidelines and refer to Figure 3-
Transmitter Mounting continuedRFT9739 Field-Mount Transmitter Instruction Manual13Output Wiring Startup TroubleshootingBefore You Begin Getting Starte
14RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual15Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sen
16RFT9739 Field-Mount Transmitter Instruction ManualPower-Supply and Sensor Wiring continuedThe base of the transmitter has three ¾-inch NPT female co
Power-Supply and Sensor Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual17Output Wiring Startup TroubleshootingBefore You Begin Gett
18RFT9739 Field-Mount Transmitter Instruction ManualPower-Supply and Sensor Wiring continued4.2 Power supply and groundingPower-supply optionsThe AC t
Power-Supply and Sensor Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual19Output Wiring Startup TroubleshootingBefore You Begin Gett
Model RFT9739 Field-Mount TransmitterInstruction ManualVersion 3 TransmittersFor technical assistance, phone the Micro Motion Customer Service Departm
20RFT9739 Field-Mount Transmitter Instruction ManualPower-Supply and Sensor Wiring continuedFigure 4-4a. Grounding detail — typicalFigure 4-4b. Ground
Power-Supply and Sensor Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual21Output Wiring Startup TroubleshootingBefore You Begin Gett
22RFT9739 Field-Mount Transmitter Instruction ManualPower-Supply and Sensor Wiring continuedThe wiring procedure is the same for the sensor and transm
Power-Supply and Sensor Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual23Output Wiring Startup TroubleshootingBefore You Begin Gett
24RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual25Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sen
26RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-1. Output terminalsTable 5-1. Output wiring terminal designationsT
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual27Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
28RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedUse RFT9739 terminals 17 and 18 for the primary mA output. Use terminals 19
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual29Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
30RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continued5.4 Frequency/pulse outputThe RFT9739 frequency/pulse output represents the
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual31Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
32RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedConfiguration for constant currentApplications with high capacitance loadin
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual33Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
34RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-8. Frequency/pulse output wiring for open collector modeFigure 5-9
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual35Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
36RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedControl output in open collector modeThe RFT9739 provides current to the co
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual37Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
38RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continued5.6 Peripheral device wiringThe wiring diagrams listed in Table 5-2 illustr
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual39Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
RFT9739 Field-Mount Transmitter Instruction ManualiContents1 Before You Begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 About
40RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-15a. Wiring to FMS-3 with LEDFigure 5-15b. Wiring to FMS-3 with LC
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual41Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
42RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-17a. Wiring to AC-powered NOCFigure 5-17b. Wiring to DC-powered NO
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual43Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
44RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continued5.7 Pressure transmitter wiringThe RFT9739 accepts input signals from a pre
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual45Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
46RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-20c. Wiring to pressure transmitter — digital communications5.8 Re
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual47Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
48RFT9739 Field-Mount Transmitter Instruction ManualOutput Wiring continuedFigure 5-22. RS-485 wiring5.10 Bell 202 multidrop networkThe RFT9739 can be
Output Wiring continuedRFT9739 Field-Mount Transmitter Instruction Manual49Output Wiring Startup TroubleshootingBefore You Begin Getting Started Mount
iiRFT9739 Field-Mount Transmitter Instruction ManualContents continued5 Output Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
50RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual51Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sen
52RFT9739 Field-Mount Transmitter Instruction ManualStartup continuedAdjusting the sight windowThe sight window in the transmitter housing cover enabl
Startup continuedRFT9739 Field-Mount Transmitter Instruction Manual53Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPow
54RFT9739 Field-Mount Transmitter Instruction ManualStartup continuedCommunication configuration modeSwitch 5 on the transmitter electronics module al
Startup continuedRFT9739 Field-Mount Transmitter Instruction Manual55Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPow
56RFT9739 Field-Mount Transmitter Instruction ManualStartup continued6.3 Custody transfer event registersEvent registers are provided for security req
Startup continuedRFT9739 Field-Mount Transmitter Instruction Manual57Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPow
58RFT9739 Field-Mount Transmitter Instruction ManualStartup continued4. Zero the transmitter in any of five ways:• Press and hold the ZERO button for
Startup continuedRFT9739 Field-Mount Transmitter Instruction Manual59Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPow
RFT9739 Field-Mount Transmitter Instruction ManualiiiContents continued7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
60RFT9739 Field-Mount Transmitter Instruction ManualStartup continued6.5 Totalizer controlThe transmitter's mass totalizer and volume totalizer c
Startup continuedRFT9739 Field-Mount Transmitter Instruction Manual61Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPow
62RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual63Output Wiring Startup TroubleshootingBefore You Begin Getting Started MountingPower-Supply and Sen
64RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continuedFollow these general guidelines when troubleshooting a Micro Motion flowm
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual65Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
66RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continued7.3 Interrogation with a HART® deviceConnect a HART Communicator to the c
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual67Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
68RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continued7.4 Troubleshooting using the transmitter displayIf the transmitter has a
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual69Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
ivRFT9739 Field-Mount Transmitter Instruction ManualContents continuedTablesTable 2-1 Security modes . . . . . . . . . . . . . . . . . . . . . . . . .
70RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continuedTable 7-4. Using overrange and sensor error messagesTable 7-5. Using slug
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual71Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
72RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continuedTable 7-6. Using informational messagesMessage Condition Corrective actio
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual73Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
74RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continued7.7 Master resetUse the switches on the transmitter electronics module to
Troubleshooting continuedRFT9739 Field-Mount Transmitter Instruction Manual75Before You Begin Getting Started MountingPower-Supply and Sensor WiringOu
76RFT9739 Field-Mount Transmitter Instruction ManualTroubleshooting continued7.8 Additional information about troubleshootingFor more information abou
RFT9739 Field-Mount Transmitter Instruction Manual77AppendixA RFT9739 SpecificationsPerformance specificationsSensor model Mass flow accuracy*ELITE li
78RFT9739 Field-Mount Transmitter Instruction ManualRFT9739 Specifications continuedFunctional specificationsOutput Signals AnalogTwo independently co
RFT9739 Field-Mount Transmitter Instruction Manual79RFT9739 Specifications continuedCommunicationSwitch allows selection of preset or user-defined set
RFT9739 Field-Mount Transmitter Instruction ManualvContents continuedFiguresFigure 1-1 RFT9739 exploded view. . . . . . . . . . . . . . . . . . 2Figu
80RFT9739 Field-Mount Transmitter Instruction ManualRFT9739 Specifications continuedPressure compensationThe analog input can accept a signal from a p
RFT9739 Field-Mount Transmitter Instruction Manual81RFT9739 Specifications continuedPower supply options and fuses85 to 250 VAC, 48 to 62 Hz, 10 watts
82RFT9739 Field-Mount Transmitter Instruction ManualRFT9739 Specifications continuedHazardous area classificationsWhen properly installed with an appr
RFT9739 Field-Mount Transmitter Instruction Manual83RFT9739 Specifications continuedPhysical specificationsHousingNEMA 4X (IP65) epoxy polyester paint
84RFT9739 Field-Mount Transmitter Instruction Manual
RFT9739 Field-Mount Transmitter Instruction Manual85AppendixB Ordering InformationRFT9739 model number matrixCode Transmitter modelRFT9739 RFT9739 tra
86RFT9739 Field-Mount Transmitter Instruction ManualOrdering Information continuedMicro Motion instruction manualsSensors•ELITE® Sensor Instruction Ma
RFT9739 Field-Mount Transmitter Instruction Manual87AppendixC Theory of OperationThe flow tubes of the Coriolis mass flow sensor are driven to vibrate
88RFT9739 Field-Mount Transmitter Instruction ManualTheory of Operation continuedMass flow measurementThe vibrating motion of the flow tube, combined
RFT9739 Field-Mount Transmitter Instruction Manual89Theory of Operation continuedρo = ρs*exp[–α∆T(1 + 0.8α∆T)]where:ρo = operating densityρs= standar